Danh mục

Đối tác



Số người truy cập

505729

Đồng hồ đo điện trở vạn năng, Mega ôm mét, Fluke- Mỹ, Fluke 1507

Giá: 12.050.000đ
Đặt hàng

Thông số kỹ thuật:

Hãng sản xuất: FLUKE- MỸ

Bảo hành: 12 Tháng

AC/DC Voltage Measurements

Accuracy

Range: 600.0 V 
Resolution: 0.1 V 
50 Hz to 400 Hz ± (% of Rdg + Digits): ±(2 % + 3)

Input Impedance

3 MΩ (nominal), < 100 pF

Common mode rejection ratio (1 kΩ unbalanced)

> 60 dB at dc, 50 or 60 Hz

Overload Protection

600 V rms or dc

Earth bond resistance measurement

Range/Resolution

20.00 Ω 0.01 Ω
200.0 Ω 0.1 Ω
2000 Ω 1.0 Ω
20.00 kΩ 0.01 kΩ

Accuracy

±(1.5% + 3)

Overload Protection

2 V rms or dc

Open Circuit Test Voltage

> 4.0 V, < 8 V

Short circuit current

> 200.0 mA

Insulation Specifications

Measurement Range

0.01 MΩ to 2000 MΩ

Test voltages

500 V, 1000V

Test Voltage Accuracy

+ 20 %, - 0 %

Short circuit current

1 mA nominal

Auto Discharge

Discharge time < 0.5 second for C = 1 µF or less

Live circuit indicator

Inhibit test if terminal voltage > 30 V prior to initialization of test

Maximum Capacitive Load

Operable with up to 1 µF load

Measure Accuracy

50 V: ±(3 % + 5) 
100 V: ±(3 % + 5) 
250 V: ±(1.5 % + 5) 
500 V: ±(1.5 % + 5) 
1000 V: ±(1.5 % + 5) to 2000 MΩ, ±(10 % + 3) above 2000 MΩ

General Specifications

Maximum voltage applied to any terminal:

600 V ac rms or dc

Storage temperature

-40 °C to 60 °C (-40 °F to 140 °F)

Operating temperature

-20 °C to 55 °C (-4 °F to 131 °F)

Temperature Coefficient

0.05 x (specified accuracy) per °C for temperatures < 18 °C or > 28 °C (< 64 °F or > 82 °F)

Relative Humidity

0 % to 95 % @ 10 °C to 30 °C (50 °F to 86 °F)
0 % to 75 % @ 30 °C to 40 °C (86 °F to 104 °F)
0 % to 40 % @ 40 °C to 55 °C (104 °F to 131 °F)

Vibration

Random, 2 g, 5-500 Hz per MIL-PRF-28800F, Class 2 instrument

Shock

1 meter drop per IEC 61010-1 2nd Edition (1 meter drop test, six sides, oak floor)

Electromagnetic compatibility

In an RF field of 3 V/M, accuracy = specified accuracy (EN 61326-1:1997)

Safety

Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-C22.2 NO. 61010-1-04, and IEC/EN 61010-1 2nd Edition for measurement category IV 600 V (CAT IV)

Certifications

CSA per standard CSA/CAN C22.2 No. 61010.1-04; TUV per standard IEC/EN 61010-1 2nd Edition

Batteries

Four AA batteries (NEDA 15A or IEC LR6)

Battery life

Insulation test use: Tester can perform at least 1000 insulation tests with fresh alkaline batteries at room temperature. These are standard tests of 1000 V into 1 MΩ with a duty cycle of 5 seconds on and 25 seconds off. 

Resistance measurements: Tester can perform at least 2500 earth bond resistance measurements with fresh alkaline batteries at room temperature. These are standard tests of 1 Ω with a duty cycle of 5 seconds on and 25 seconds off.

Size

5.0 cm H x 10.0 cm W x 20.3 cm L (1.97 in H x 3.94 in W x 8.00 in L)

Weight

550 g (1.2 lb)

IP Rating

IP40

Altitude

Operating: 2000 m CAT IV 600 V, 3000 m CAT III 600 V 
Non-operating (storage): 12,000 m

Over-range capability

110% of range

Included accessories

TL224 Test Leads, TP74 Test Probes, clips PN 1958654 (red) and PN 1958646 (black), holster and remote probe

Includes

Remote probe

Test leads

Test probes

Alligator clips

User documentation