Máy đo pH,ORP, nhiệt độ để bàn HANNA- Mỹ, HI 2223, –2.00 – 16.00 pH, -2.000 – 16.000 pH
Thông số kỹ thuật:
Hãng sản xuất: Hanna- Mỹ
Xuất xứ: Romania
Bảo hành: 12 Tháng đối với thân máy
6 Tháng đối với đầu điện cực
|
Dải đo |
pH |
–2.00 to 16.00 pH; –2.000 to 16.000 pH |
|
mV |
±999.9 mV; ±2000 mV |
|
|
Temperature |
– 20.0 to 120.0 °C |
|
|
Độ phân giải |
pH |
0.01 pH; 0.001 pH |
|
mV |
0.1mV(± 999.9mV); 1mV(± 2000mV) |
|
|
Temperature |
0.1 °C |
|
|
Cấp chính xác |
pH |
±0.01 pH; ±0.002 pH |
|
mV |
±0.2mV(± 999.9mV); |
|
|
Temperature |
± 0.2 °C (excluding probe error) |
|
|
Hiệu chuẩn pH |
Up to 5 point calibration, 7 standard buffers available (1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45) |
|
|
Bù nhiệt độ |
Manual or automatic from -20 to 120°C |
|
|
Đầu đo nhiệt độ |
HI 1131P glass body, single junction, refillable with BNC connection and 1 m (3.3′) Cable HI 7662 stainless steel probe |
|
|
Kết nối PC |
USB |
|
|
Ghi dữ liệu |
500 points |
|
|
Kích thước |
235 x 222 x 109 mm (9.2 x 8.7 x 4.3″) |
|
|
Trọng lượng |
1.3 kg (2.9 lb.) |
|

- pH Calibration Check™
- On-screen electrode condition and response time
- Up to five point calibration with seven standard buffers
- Diagnostic alerts through icons
- Automatic Temperature Compensation
- Log up to 500 samples
- GLP features
- PC interface via US
A properly manufactured and maintained pH electrode will retain its measuring integrity for a long time. As a result of many factors such as age, poor maintenance or improper handling, any electrode in time will lose its integrity.
The most common cause for pH measurement inaccuracies is an unclean or improperly cleaned electrode. This is very important to note, because during calibration, the instrument assumes that the electrode is clean and that the standardization curve created during the calibration process will remain a valid reference until the next calibration.
A dirty electrode or wrong calibration approach can contribute to buffer solution contamination. A contaminated buffer solution can present a major problem during calibration due to the fact that it is considered the only reference.
The HI 2223 is a pH benchtop meter featuring our exclusive Calibration Check™ diagnostics for both pH?electrodes and buffer solutions during calibration. This instrument compares the characteristics of the pH electrode from one calibration to the next. In the case of large variances in the electrode condition, it alerts the user that the electrode needs to be properly cleaned prior to calibration and measuring.
The second feature is to detect if the calibration buffer solution is contaminated. After calibration, the probe condition is evaluated and an indicator is displayed informing the user of the overall pH electrode status.
This meter can measure using ORP electrodes (pH channel input), due to it’s capability to measure mV with a resolution up to 0.1 mV.
HI 2223 also features five point calibration with seven standard buffers, Automatic Temperature Compensation and GLP capabilities that allow for the storage and retrieval of all data regarding pH. With a built-in logging function, measurements are stored in non volatile memory and can be transferred to a PC through the USB port. Users can manually log up to 200 records and interval log up to 500 records.
Order Information:
HI 2223-01 (115V) and HI 2223-02 (230V) are supplied with HI 1131B pH electrode, HI 7662 temperature probe, HI 76404N electrode holder, HI 70004 pH 4.01 buffer solution sachet, HI 70007 pH 7.01 buffer solution sachet, HI 700661 cleaning solution sachet, HI 7071S electrolyte solution (30 mL), 12 VDC adapter and instructions.