Thiết bị đo LCR để bàn IETLABS- Mỹ , 7600 Plus

Liên hệ
Mã sản phẩm: 158
Liên hệ trực tiếp

Thông tin cơ bản:

Hãng sản xuất: IETLABS- Mỹ

Xuất xứ: Mỹ

Bảo hành: 12 Tháng

Thông số kỹ thuật:

Tham số đo

Dải đo

Cấp chính xác

Nhanh

Trung bình

Chậm

Cs, Cp 

00000 .01 fF to 9.999999 F 

±0.5% 

±0.25% 

±0.05% 

Ls, Lp 

0000.001 nH to 99.99999 H 

±0.5% 

±0.25% 

±0.05% 

.0000001 to 99 .99999 

±0.005 

±0.0025 

±0.0005 

.0000001 to 999999.9 

±0.005 

±0.0025 

±0.0005 

Z. Rs, Rp, ESR,Xs 

000.0001 mΩ to 99 .99999 MΩ

±0.5% 

±0.25% 

±0.05% 

Y, Gp, Bp

00000 .01 µS to 9.999999 MS 

±0.5% 

±0.25% 

±0.05% 

Góc pha 

-180.0000° to +179.9999° 

±1.8° 

±0.9° 

±0.18° 

 

Test Tần số: Dải đo: 10 Hz to 2 MHz;  Độ phân giải liên tục: 0.1 Hz from 10 Hz to 10 kHz, 5 digits>10kHz;  Cấp chính xác: +/-(0.01% + 0.10Hz)
Tốc độ đo: Nhanh: 120 meas/sec;  Trung bình : 16 meas/sec – 8 meas/sec below 150kHz Chậm: 2 meas/sec – 1 meas/sec below 150kHz
Ranging: Automatic, Range Hold or user selectable
Trigger: Internal (automatic), External (RS-232, IEEE-488.2 or Handler interfaces) and Manual
AC Test Signal:
Điện áp: 20 mV to 5.0 V (open circuit) up to 500kHz in 5 mV steps 20 mV to 1.0 V (open circuit) 500kHz-1MHz in 5 mV steps 20 mV to 0.5 V (open circuit) >1MHz in 5 mV steps 
Dòng điện: 250 μA to 100 mA (short circuit) in 50 μA steps Max. Compliance 3V < 500kHz.
Trở kháng nguồn: 25Ω, 400Ω, 6.4kΩ, or 100kΩ, range dependent
DC Bias Voltage: Internal: 2.0 V External: 0 to +/-200V
Sweep Result: 
Primary parameter vs. frequency, voltage or current Graphical or Tabular Format Up to 200 measurement points per sweep
Sequencing Result: Displays up to 6 sequential test results, primary and/or secondary
AutoAcc: Automatic calculation and display of overall instrument accuracy for selected settings, test conditions, and device under test
Interfaces: Standard: USB Host Port, RS-232, Handler, Printer Port Optional: IEEE-488.2 Adapter to RS-232
Charged Capacitor Protection: √8/C for Vmax < 250 V; √2/C for Vmax < 1000V C = Capacitance in farads of the device under test
Measurement Delay: Programmable from 0 – 1000 ms in 1 ms steps
Averaging: Programmable from 1 – 1000 Median value mode
Data Storage: USB Host Port USB1.1 Compliant, ASCII format
Program Storage: Memory internal, USB Host Port, ASCII format
Mechanical: 
Bench mount with tilt bail Dimensions: (w x h x d): 16 x 6 x 14in (410 x 150 x 360mm) Weight: 17 lbs (8kg) net, 23 lbs (10.5kg) shipping
Safety: 
IEC61010-1: 2001 CAT 1, Pollution Degree 2
EMC: 89/336/EEC, 92/31/EEC, 93/68/EEC

IET/QuadTech 7600 Plus Precision LCR Meter

  • Measures 14 Impedance Parameters (2 simultaneously)
  • 0.05% Basic Measurement Accuracy
  • Programmable Frequencies from 10Hz to 2MHz
  • AC Test Voltage Programmable from 20mV to 5V
  • DC Bias Voltage, 2V Internal or 0 to 200V External
  • Fast Measurement Speeds up to 120/Second
  • Automatic Open/Short Zeroing Eliminates Connection Errors
  • High Resolution LCD with 7 Digit Display
  • 14 Hi/Lo Sorting Bins for Pass/Fail Testing
  • RS-232, USB, Printer, PLC Interfaces all Standard, (IEEE-488 Adapter optional)
  • Results Displayed as Deviation or %Deviation from a Nominal Value
  • Internal Storage and Recall of Test Setups
  • Constant Voltage Maintained to the Device Under Test
  • Ability to Automatically Perform 6 Unique Tests in Sequence
  • Keypad Lockout Prevents Changes to the Test Setup
  • On site calibration with 7000-09 Calibration Kit
  • Programmable Average and Delay Times

7600 Plus Precision LCR Meter Basic Specifications

(See datasheet for full specifications)

 

Measured Parameters and Basic Measurement Accuracy*: Any two of 14 parameters measured and displayed simultaneously, user selectable

Parameter  Measurement Range  Fast  Medium  Slow 
Cs, Cp  00000 .01 fF to 9.999999 F  ±0.5%  ±0.25%  ±0.05% 
Ls, Lp  0000.001 nH to 99.99999 H  ±0.5%  ±0.25%  ±0.05% 
.0000001 to 99 .99999  ±0.005  ±0.0025  ±0.0005 
.0000001 to 999999.9  ±0.005  ±0.0025  ±0.0005 
Z. Rs, Rp, ESR,Xs  000.0001 mΩ to 99 .99999 MΩ ±0.5%  ±0.25%  ±0.05% 
Y, Gp, Bp 00000 .01 µS to 9.999999 MS  ±0.5%  ±0.25%  ±0.05% 
Phase Angle  -180.0000° to +179.9999°  ±1.8°  ±0.9°  ±0.18° 

*At optimum test signal levels, frequencies, DUT values and without calibration uncertainty. Capacitance (Cs/Cp), Inductance (Ls/Lp), Resistance (Rs/Rp), Dissipation (D) and Quality (Q) Factors, Impedance Z , Admittance Y , Phase Angle (0-), Equivalent Series Resistance (ESR), Conductance (Gp), Reactance (Xs), Susceptance (Bp) Note: s = series, p = parallel, ESR equivalent to Rs
Test Frequency: Range: 10 Hz to 2 MHz, continuous Resolution: 0.1 Hz from 10 Hz to 10 kHz, 5 digits>10kHz Accuracy: +/-(0.01% + 0.10Hz)
Measurement Speed: Fast: 120 meas/sec Medium: 16 meas/sec – 8 meas/sec below 150kHz Slow: 2 meas/sec – 1 meas/sec below 150kHz
Ranging: Automatic, Range Hold or user selectable
Trigger: Internal (automatic), External (RS-232, IEEE-488.2 or Handler interfaces) and Manual
AC Test Signal:
Voltage
: 20 mV to 5.0 V (open circuit) up to 500kHz in 5 mV steps 20 mV to 1.0 V (open circuit) 500kHz-1MHz in 5 mV steps 20 mV to 0.5 V (open circuit) >1MHz in 5 mV steps 
Current: 250 μA to 100 mA (short circuit) in 50 μA steps Max. Compliance 3V < 500kHz.
Source Impedance: 25Ω, 400Ω, 6.4kΩ, or 100kΩ, range dependent
DC Bias Voltage: Internal: 2.0 V External: 0 to +/-200V
Sweep Result: Primary parameter vs. frequency, voltage or current Graphical or Tabular Format Up to 200 measurement points per sweep
Sequencing Result: Displays up to 6 sequential test results, primary and/or secondary
AutoAcc: Automatic calculation and display of overall instrument accuracy for selected settings, test conditions, and device under test
Interfaces: Standard: USB Host Port, RS-232, Handler, Printer Port Optional: IEEE-488.2 Adapter to RS-232
Charged Capacitor Protection: √8/C for Vmax < 250 V; √2/C for Vmax < 1000V C = Capacitance in farads of the device under test
Measurement Delay: Programmable from 0 – 1000 ms in 1 ms steps
Averaging: Programmable from 1 – 1000 Median value mode
Data Storage: USB Host Port USB1.1 Compliant, ASCII format
Program Storage: Memory internal, USB Host Port, ASCII format
Mechanical: Bench mount with tilt bail Dimensions: (w x h x d): 16 x 6 x 14in (410 x 150 x 360mm) Weight: 17 lbs (8kg) net, 23 lbs (10.5kg) shipping
Safety: IEC61010-1: 2001 CAT 1, Pollution Degree 2
EMC: 89/336/EEC, 92/31/EEC, 93/68/EEC